18

EL232 XCPD Installation Manual

3.3.1

Loss of Signal Display

Press the Reset switch to enter new values into the
E L 2 3 2 X P C D , o r t u r n o ff t h e p ow e r, o r wa i t
approximately 30 seconds and the EL232 XPCD will
reset itself.

The following tables illustrate switch settings for the
loss of signal display.

3.4

Serial Data Output

The EL232 XPCD comes standard with an active 20
mA current loop transmitter which echoes the serial
data received by the EL232 XPCD. This data is
echoed as a 20 mA loop regardless of whether the
input source was RS232 or current loop.

The output is active and supplies current from a
voltage source of approximately 8 VDC. The current
transmitter is an HP-4100 and can transmit data
cleanly at 9600 baud. The HP-4100 supplies mark
currents of >12 mA and space currents of <2 mA.

Certain scale industry remote displays and printers use
darlington type optocouplers as input devices.
Unbiased 4N33 type opto isolators can be turned on
by 2 mA. These are not true current loop receivers. If
a device which the EL232 XPCD is transmitting to
cannot properly receive the data, change the 4N33 to a
4N25 opto isolators and the problem will normally
disappear.

Figure 3-10 illustrates how to connect the current loop
output of the EL232 XPCD.

Figure 3-10. 20 mA Active Current Loop Output

You can check the current loop output with a simple
jumper connected between pin J2-3and J2-4. The
LED on the CPU should flash on and off when data is
transmitted.

Note:

When using a passive current loop input, the serial

output is not available. Consult the factory for additional
information.

Indicator

Indicator Type

SW1-3

SW1-4

SW1-5

SW1-6

Condec

On

On

On

On

General Freedom

Off

On

On

On

Electro 560

On

Off

On

On

A n D

Off

Off

On

On

WI-110, WI-120

On

On

Off

On

Toledo

Off

On

Off

On

Flexweigh

On

Off

Off

On

Analogic

Off

Off

Off

On

Masstron

On

On

On

Off

Fairbanks

Off

On

On

Off

Streeter Richardson

On

Off

On

Off

Cardinal 738

Off

Off

On

Off

Generic

Off

Off

Off

Off

Table 3-3. Indicator Switch Settings

Loss of Signal Display

Display

SW1-7

Last data

Off

- - - - - -

On

Table 3-4. Loss of Signal Display

Input Signal Type

Interface

SW1-8

RS232

On

20 mA

Off

Table 3-5. Input Signal Type

GND

PRINT

K4

K3

K2

K1

+5V

CL4 IN-

CL4 IN+

CL3 IN-

CL3 IN+

CL2 IN-

CL2 IN+

CL1 IN (-)

CL(+) OUT

J2

CL(-) OUT

CL1 IN (+)

RS-232 (+)

RS-232 (-)