Internal count, Calibration procedure, Used to switch between internal count and a/d – Rice Lake Digi DI-28SS Bench Scale User Manual

Page 16: Raw data

Advertising
background image

6.2. Internal Count
This function is provided to check the internal count value from
the A/D (pre-amp)
To Enter Internal AD Count Mode

1. Press and hold down the [REZERO] key.
2. While holding , press the [

DIGIT] key two times, the

[TARE] key once.
3. Then release the [REZERO] key.

In This Mode, Each Key Functions As Follows

[

↑↑↑↑ INCREASE] key

Used to switch between internal
count and A/D

Raw Data

[REZERO] key

Used to rezero the internal counts

[TARE] key

Used to exit from internal count mode


















[14]


6.5.2. Calibration Procedure

PROCEDURE KEY

OPERATION

DISPLAY

Remarks

1. Enter calibration
mode.

[REZERO]
[

DIGIT] [TARE]

[TARE]

888888

Press and hold
[REZERO] key
while pressing
[DIGIT], [TARE],

| 1

[TARE] then release
[REZERO] key.

2. Enter the minimum
display.

[

↑↑↑↑ INCR]

| 2

Press [

↑↑↑↑ INCR] key.

(Ex. 0.002)

[REZERO]

[ 00.000

Press [REZERO] to
advance through
calibration

3. Enter the capacity
weight.

[

] [ ↑↑↑↑ ] [ ↑↑↑↑ ]

[

↑↑↑↑ ] [ ↑↑↑↑ ] [ ↑↑↑↑ ]

[

↑↑↑↑ ]

[ 06.000

Press [DIGIT] once
and [

↑↑↑↑ INCR] key six

times. (Ex. 6KG)

[REZERO]

〈 06.000

Press [REZERO] to
advance through
calibration

4. Enter weight to be
used for calibration. (If
other than capacity.)

[

] [ ↑↑↑↑ ] [ ↑↑↑↑ ]

[

↑↑↑↑ ] [ ↑↑↑↑ ] [ ↑↑↑↑ ]

[

↑↑↑↑ ]

〈 02.000

Press [DIGIT] once
and [

↑↑↑↑ INCR] key six

times. (Ex. 2KG)

[REZERO]

CAL 0

Press [REZERO] to
advance through
calibration

5. Set zero point.

[REZERO]

CAL SP

6. Place weight on
platter from step 4.

[REZERO]

xxxxxx

When calibration is
complete scale shows
internal count.

7. Exit internal count
mode.

[TARE]

xx.xxx

Press [TARE] key to
return to weighing
mode.

8. Rezero scale place
weight on scale to test
calibration.

[REZERO]

[19]

Advertising