6.5.1. Span Adjustment
This Mode Is Provided To Calibrate Scale

Accessing Calibration Mode
1. Press and hold down the [REZERO] key.
2. While holding , press the [

DIGIT] key, the [TARE] key, the

[TARE] key.
3. Then release the [REZERO] key.

In This Mode, Each Key Functions As Follows

[

↑↑↑↑ INCREASE] key Used to increase value of digit

[

DIGIT] key

Used to change digits as needed

[REZERO] key

Used to save changes and advance through
calibration

[TARE] key

Used to exit calibration mode without
saving changes

















[18]


6.3. Set Up And Calibration

6.3.1. Specification Change
Function key for spec. change.

Accessing Operational Specs. (50 – 59)
1. Press and hold down the [REZERO] key.
2. While holding , press the [

DIGIT] key three times .

3. Then release the [REZERO] key.

In This Mode, Each Key Functions As Follows

[

DIGIT] key

Used to input 0

[

↑↑↑↑ INCREASE] key

Used to input 1

[REZERO] key

Used to save the specification and increment to
the next specification

[TARE] key

Used to exit from specification mode


THE SLEEP TIMER IS LOCATED IN OPERATIONAL SPECS.
Spec.50 = sleep timer ( 0000 = none to 1111 = 15 min. )

6.3.2. Specification Change (Continued)
Function key for spec. change.

Accessing Weights & Measures Specs. (00 – 09 / 20 – 29)
1. Press and hold down the [REZERO] key.
2. While holding , press the [

DIGIT] key, the [TARE] key, the[

DIGIT] key.
3. Then release the [REZERO] key.

In This Mode, Each Key Functions As Follows

[

DIGIT] key

Used to input 0

[

↑↑↑↑ INCREASE] key Used to input 1

[REZERO] key

Used to save the specification and increment to
the next specification

[TARE] key

Used to exit from specification mode

[15]