IQ plus 2100 Installation Manual

3.2

Menu Structures and Parameter Descriptions

The following sections provide graphic representations of the IQ plus 210 menu structures. In the actual menu
structure, the values under each parameter are arranged horizontally. To save page space, menu choices are shown in
vertical columns. The factory default setting appears at the top of each column.

Most menu diagrams are accompanied by a table that describes all parameters and parameter values associated with
that menu. Default parameter values are shown in bold type.

3.2.1

Configuration Menu

Figure 3-6. Configuration Menu

CONFIG Menu

Parameter

Choices

Description

Level 2 submenus

GRADS

5000
number

Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1–10 000 and should be consistent with legal requirements and environmental
limits on system resolution.

To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBN

OFF
0.5D
1D
3D

Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZRANGE parameter). Selections are ±
display divisions. Maximum legal value varies depending on local regulations.

ZRANGE

100%
1.9%

Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is
± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be at
standstill to zero the scale. Use 1.9% for legal-for-trade applications.

MOTBAN

1D
2D
3D
5D
10D
20D
OFF

Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected for 1 second or more, the standstill symbol lights. Some
operations, including print, tare, and zero, require the scale to be at standstill. Maximum
legal value varies depending on local regulations.

If OFF is selected, ZTRKBN should also be set to OFF.

Table 3-2. Configuration Menu Parameters

XXXXXXX

XXXXXXX

GRADS

5000

OFF

1D

0.5D

3D

ZTRKBN

ZRANGE

100%

1.9%

OFF

2D

1D

3D

MOTBAN

10D

5D

20D

FS+2%

FS+9D

FS+1D

FS

OVRLOA

1

8

4

16

FILTER

8RT

4RT

16RT

number

VERS

DEFLT

DIG IN

SERIAL

CALIBR

CONFIG

FORMAT