IQ plus 2100 Installation Manual

4.0

Calibration

The IQ plus 210 can be calibrated using the front panel, EDP commands, or the Revolution

configuration utility.

Each method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in setup mode (display

reads

CONFIG

) and remove all weight from

the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.

2. Press the

RIGHT

(

UNITS

) button until the

display reads

CALIBR

(see Figure 4-1). Press

ENTER

(

ZERO

) to go to zero calibration

(

WZERO

).

3. With

WZERO

displayed, press

ENTER

to

calibrate zero. The indicator displays

°CAL°

while calibration is in progress. When
complete, the A/D count for the zero
calibration is displayed. Press

ENTER

again to

save the zero calibration value and return to the

WZERO

prompt or use the procedure shown in

Figure 4-2 to edit the value. When done, press

RIGHT

to go to the

WVAL

prompt.

4. With

WVAL

displayed, press

ENTER

to show

the test weight value. If the value is equal to the
test weight you are using, press

ENTER

again

to save the value and return to the

WVAL

prompt; if the value is incorrect, use the
procedure shown in Figure 4-2 to edit the
value. When done, press

RIGHT

to go to the

WSPAN

prompt.

NOTE:

When calibrating the indicator with

LB/OZ as the primary unit, enter the WVAL value
in ounces. For example, to calibrate a 5 lb scale
using LB/OZ, enter 80 (5 lb x 16 oz/lb) as the
WVAL value.

5. With

WSPAN

displayed, place test weights on

the scale then press

ENTER

to calibrate span.

The indicator displays

°CAL°

while calibration

is in progress. When complete, the A/D count
for the span calibration is displayed. Press

ENTER

again to save the span calibration value

and return to the

WSPAN

prompt or use the

procedure shown in Figure 4-2 to edit the
value. When done, press

RIGHT

to go to the

REZERO

prompt.

Figure 4-2. Editing Procedure for Numeric Values

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration

A/D count value

WSPAN

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO

XXXXXXX

XXXXXXX

VERS

DEFLT

DIG IN

SERIAL

CALIBR

CONFIG

FORMAT

To edit numeric values, press the

UNITS (RIGHT)

button to select the leftmost digit. Each digit flashes
when selected: Press

ZERO (ENTER)

to

increment the value of the selected digit; press the

RIGHT

button to move right to the next digit.

Once the last digit has been edited, press

RIGHT

again and the digit stops flashing. Press

ENTER

to save the value entered and return to the menu level
above, or press

RIGHT

to edit the value again,

starting with the leftmost digit.