1 front panel calibration, Front panel calibration – Rice Lake RailBoss™ Rail Scales - Installation Manual User Manual

Page 17

Advertising
background image

Calibration

13

5.3.1

Front Panel Calibration

The CALIBR submenu (under the SCALES menu, see
Figure 5-6) is used to calibrate the 920i. The zero,
span, and linear calibration point displays provide a set
of softkeys used specifically for calibration procedures:

+/–

Toggles to allow entry of negative or
positive values

Last Zero

Recalls the last established zero value to
allow calibration without removing tests
weights or product from scale.

Calibrate

Performs calibration for the selected
point

Temp Zero

Temporarily zeroes the displayed weight
o f a n o n - e m p t y s c a l e . A f t e r s p a n
calibration, the difference between the
temp zero and the previously calibrated
zero value is used as an offset.

Millivolts

(or

Counts

)

Toggles between display of captured A/
D counts and captured millivolts values;
allows entry of calibration values in mV
or counts

03/01/2011

01:57PM

Calibrate

WZERO

WVAL

REZERO

1291

Deadload Calibration

+/-

Millivolts

Zero Count =>

Last Zero

Temp Zero

CALIBR

SCALE1

SCALES

Figure 5-6. WZERO Calibration Display

To calibrate the indicator using the front panel, do the
following:

1. Ensure the J-Box trimming procedure has been

successfully completed.

2. Place the indicator in setup mode by removing

t h e l a rg e f i l l i s t e r- h e a d s c r e w f r o m t h e
enclosure. Insert a screwdriver or similar tool
into the access hole and press the setup switch
once. The indicator display changes to show
scale configuration menus.

3. With the

SCALES

menu highlighted, press the

down

key, then select the Scale 1. Press

down

again (

GRADS

parameter highlighted), then

press

left

to highlight the

CALIBR

submenu (see

F i g u r e 5 - 5 ) . P r e s s

d o w n

t o g o t o z e r o

calibration (

WZERO

). See Figure 5-6.

4. Ensure scale reading is at zero, then press

down

again to show the current WZERO value, then
press the

Calibrate

softkey to calibrate zero.

When complete, the new A/D count for the
zero calibration is displayed. Press

Enter

to

save the zero calibration value and go to the
next prompt (

WVAL

).

5. With

WVAL

displayed, press

down

to show the

stored calibration weight value. Use the
numeric keypad to enter the actual value of the
RailBoss Calibrator load to be used (30,000
lbs. is recommended), then press

Enter

to save

the value and go to span calibration (

WSPAN

).

6. Use the RailBoss Calibrator to place a 30,000

lb. load on the scale. Press

down

again to show

the current WSPAN value, then press the

Calibrate

softkey to calibrate span. When

complete, the new A/D count for the span
calibration is displayed. Press

Enter

again to

save the span calibration value and go to the
next prompt (

WLIN

).

7. When complete, press the

Save and Exit

softkey

to exit setup mode.

Advertising