14

Railboss Railroad Track Scale

5.3.2

iRev Calibration

The iRev 4 Calibration Wizard provides step-by-step
scale calibration. With the 920i connected to a PC,
select the Calibration Wizard from the Tools menu on
the iRev 4 Scales display, then follow the steps listed
below to calibrate the scale.

1. On the first Calibration Wizard display (see

Figure 5-7), select Standard Calibration. Click
the

Next

button to continue.

Figure 5-7. iRev 4 Calibration Wizard

2. Select the scale to calibrate (Scale #1).

Figure 5-8. iRev 4 Scale Selection Display

3. Enter the RailBoss Calibrator load value used

to calibrate the scale (30,000 lb).

Figure 5-9. iRev 4 Test Weight Value Display

4. Ensure the scale load is at zero. Press the

Calibrate Zero

button to perform the zero

calibration. A message appears when the
process is complete.

Figure 5-10. iRev 4 Zero Calibration Display

5. Using the RailBoss Calibrator, apply the 30000

load to the scale (see Section 5.1). Press the

Calibrate Span

button to perform the span

calibration. A message appears when the
process is complete.

Figure 5-11. iRev 4 Span Calibration Display

6. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.