3 accessing the audit trail – Rice Lake TradeRoute HL Series - M2000 Scale User Manual

Page 26

Advertising
background image

22

the audit trail. The audit trail is permanent and cannot be disabled or erased
by removing the internal battery inside the indicator. The Audit Trail has two

counters, the Calibration counter and the Parameter counter. Changing
parameters that affect the calibration of weight will increment the Calibration
counter by 1. All other parameters will increment the Parameter counter.

The counters will count from 000 to 999 before rolling over again. Changing
one or more parameters will only increment the counter by 1 per calibration

session. Only an actual change of value to the calibration parameter will
register to the audit trail.

Important Note:

Because the audit trail becomes active at the factory, the counters may not show 0
even when the indicator is new out of the box.

5.3.3 Accessing the Audit Trail

The audit trail can only be accessed from normal weighing mode and cannot
be accessed while you are in calibration mode. Enter 1000 followed by the

PRINT SELECT. The message “Audit” will briefly be displayed on the display.
Shortly after the indicator will cycles through 3 times displaying the date of
when the last change was made to the indicators calibration parameters, the

calibration (CAL) counter and the configuration (CFG) counter. You may
press the CLEAR key at any time to cancel this operation.