9 serial communications parameters – Rice Lake TradeRoute HL Series - M2000 Scale User Manual

Page 40

Advertising
background image

36

This function will blink the display every time a key is pressed. This gives a
sense of feedback to the end user when a key is pressed. This is especially

handy in noisy environments where the operator may not be able to here the
key beeps from the indicator. Parameters are 0 (off) or 1 (on) followed by the
[TARE] key. Error 32 will be displayed for an invalid entry.

80 Set Time
Time is set out side of calibration mode in normal weighing mode (you must

exit cal mode). This parameter allows you to set the battery backed real time
clock on the indicator. Pressing 80 followed by the [PRINT/SELECT] key will
display the current time on the display. To change the time, enter your new

time as HHMMSS followed by the [TARE] key. Example: to change the time to
9 hours 33 minutes and 30 seconds enter 093330 followed by the [TARE] key.
81 Set Date

Date is set out side of calibration mode in normal weighing mode. This
parameter allows you to change the battery-backed date on the indicator.
Pressing 81 followed by the [PRINT/SELECT] key will display the current date

on the display. To change the time, enter your new time as YYMMDD followed
by the [TARE] key. Example: to change the date to December 14 year 2001
enter 011214 followed by the [TARE] key.

8.9 SERIAL COMMUNICATIONS Parameters

30 baud rate for COM 1 (6)
Changes the baud rate for COM1. Valid parameters are 0-9. An invalid entry

will display error 52.
32 baud rate for COM 2 (6)
Changes the baud rate for COM1. Valid parameters are 0-9. An invalid entry

will display error 52.
0 : 150
1: 300

2: 600
3: 1200

4: 2400
5: 4800
6: 9600 (factory default)

7: 14400
8: 19200
9: 32400

9600 baud is the default baud rate for both ports. Note: only a single COM port
can be used for printing.


31 Parity for COM 1 (0)
Changes the parity setting for COM1. Valid parameters are 0-2. An invalid

entry will display error 53.