6 indicator power up parameters – Rice Lake TradeRoute HL Series - M2000 Scale User Manual

Page 38

Advertising
background image

34

Before spanning the indicator for the first time the indicator must be
deadloaded. Ensure that the scale has all test weights removed from the scale

and the scale is stable before performing a deadload. To perform the deadload
function press 1 followed by the [TARE] key. The indicator will perform the
deadload function. After the deadload has completed the indicator will briefly

display the deadload value in AD raw counts before returning to weight display
mode.

13 Set Span
This command is used to span the indicator with a known test weight. Place
the test weights on the scale. When the set span function has been activated

the current displayed weight will be frozen on the display. Using the numeric
keypad enter the known test weight followed by [TARE]. The display should
now show the new weight on the scale. You can cancel a span any time by

pressing the [CLEAR] button before [TARE] is pressed.

99 Exit calibration mode

This parameter will exit calibration mode and restart the indicator. All
calibration changes are stored to flash memory and the audit trail is updated.

8.6 INDICATOR POWER UP Parameters

6 Power On Units (0)
Determines what units the indicator defaults to when the indicator is turned
on. Entering 1 followed by the [tare] key selects Kilograms. Entering 0 selects

Pounds. An invalid entry will display Error 13.
9 Power ON Zero Scale Message (0)
With this parameter set to 1 the indicator will power up displaying “Zero” on

the display. The operator must Zero the scale before continuing. This must be
set for certain legal for trade requirements. If the parameter is set to 0 the
indicator powers up displaying weight. An invalid parameter entry will display

error 28.
10 Power Switch Bypass (1)

This parameter controls the keypad ON/OFF switch for the indicator. Entering
1 for this parameter disables the ON/OFF switch and the indicator can never
be turned off from the keypad. Also when the indicator power is plugged into

the wall the indicator will turn on without pressing the ON button. This should
be used in applications such as control systems where the indicator must
power up running after a power outage. An invalid entry will display error 56.

46 Power up channel select (1)
This parameter will determine what channel to be the default channel at boot
up. This parameter will only be implemented if the calibration values meet the

criteria.



1 channel 1 as default at startup



2 channel 2 as default at startup (if enabled)



3 channel 3 as default at startup (if enabled)

Advertising