Troubleshooting • 109

English

If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 109.

CAUTION!

Bodily injury

If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 109, you may not use the device.
– Have servicing and repairs carried out

exclusively by seca service or by an
authorized service partner.

– Please see the section entitled

"Servicing/recalibration" on page 111.

10.TROUBLESHOOTING

Problem

Cause / Solution

No weight displayed
with a load on the scale.

The scale has no power supply.

- Check whether the scale is switched on.

0.0 does not appear
before weighing.

A load was already on the scale before it was
switched on.

- Remove the load from the scale.
- Switch the scale off and back on again.

One segment is illumi-
nated either continu-
ously or not at all.

The corresponding point has a fault.

- Inform the service department.

The display

appears.

Battery block power is running down.

- Recharge battery block as soon as possible.

The display "bAtt"
appears.

Battery block discharged.

- Charge battery block.

The display "StOP"
appears.

Maximum load has been exceeded.

- Remove the load from the scale.

The display "tEMP"
appears.

The ambient temperature of the scale is too high
or too low.

- Set up the scale in an ambient temperature

between +10 °C and +40 °C.

- Wait for around 15 minutes until the device

has adapted to the ambient temperature.