2 checking the content of the calibration counter – Seca 645 band_1 User Manual

Page 112

Advertising
background image

112 •

11.2 Checking the content of the calibration counter

This seca scale is calibrated. Calibrations may only be
performed by authorized agencies. To guarantee this,
the scale is equipped with a calibration counter to
record any change in calibration-related data.

If you want to check whether the scale has been pro-
perly calibrated, follow the instructions below.

1. If necessary, turn off the scale.

2. Press and hold any key and start the scale.

The current content of the calibration counter flas-
hes in the display for a few seconds.

3. Compare the content of the calibration counter out-

put with the number given on the calibration coun-
ter sticker.

Both numbers have to match for the calibration to be
valid. If the sticker and the calibration counter do not
match, the scale must be recalibrated. Please contact
your service partner or seca service. Once the scale has
been recalibrated, a new, updated calibration counter
sticker is used to identify the status of the calibration
counter. The person authorized to perform the recalib-
ration secures this sticker with an additional seal. The
calibration counter sticker can be ordered from seca
service.

Advertising