After switching on for
the first time, measure-
ment results are trans-
mitted and two beeps
are heard.

• The device was unable to send measurement

results to the wireless receiver (seca wireless
printer or PC with seca USB wireless adapter).
- Ensure that the scale is integrated in the

wireless network.

- Ensure that the receiver is switched on.

• Reception is being interfered with by HF equip-

ment (e.g. cell phones) in the vicinity.
- Keep HF equipment at least 1 meter away

from transmitters and receivers in the seca
wireless network. The actual transmission
output of HF equipment may require mini-
mum distances of more than 1 meter. For
details, go to www.seca.com.

NOTE
If this fault is not eliminated, there will be no
fresh acoustic warning when further
attempts are made to send data.

Only the "SYS" item is
visible in the "rF" menu.

• The wireless module is deactivated.

- Activate wireless module (see "Activating the

wireless module (SYS)" on page 103).

Only the "SYS" and "Lrn"
items are visible in the
"rF" menu.

• The wireless module is activated and no

wireless group is set up.
- Set up wireless group (see "Setting up a

wireless group (Lrn)" on page 103).

The items "APrt" and
"Time" are not visible in
the "rF" menu.

• No wireless printer registered to the wireless

group.
- Register the wireless printer to the wireless

group via the "Lrn" menu item (see "Setting
up a wireless group (Lrn)" on page 103).

After calling up the
menu, the "rF" item is
not displayed.

• The scale’s wireless module is defective.

- Inform seca service.

The display "Er:X:11"
appears.

The scale has too high a load or too high a load in
one corner.

- Take the load off the scale or distribute the

weight more evenly.

- Restart the scale.

The display "Er:X:12"
appears.

The scale has been switched on with too high a
load.

- Remove the load from the scale.
- Restart the scale.