Welch Allyn CP 150 spirometry option - User Manual User Manual

Page 16

Advertising
background image

4. Insert the mini USB cable connector into the spirometer sensor mini USB port.

a. Insert the USB cable into the spirometer sensor mounting bracket groove to

secure the cable.

b. Insert the USB cable connector into the device's first USB port, furthest to the

right.

Note

The mounting bracket is designed to protect the spirometer
sensor and USB cable and only accepts the USB cable mini
connector when the spirometer sensor label faces outward.

5. Verify that the spirometer sensor and pressure tubing are clean and undamaged.

Look for signs of deterioration, including but not limited to cracks, cuts, discoloration,

or oxidation. If any part exhibits any of these symptoms, replace it.
a. Attach the pressure tubing to the spirometer sensor.
b. Attach a flow transducer to the pressure tubing.

CAUTION Hand-tighten the spirometer sensor and flow
transducer connectors to avoid damaging the connectors.

The CP 150 software automatically activates the spirometry functions throughout the

software. Once the software recognizes the sensor, the Spirometry button appears

in the Content area.

12 Setup

CP 150 spirometry option