The results display for a single-flow calibration after no air has moved for three

seconds.

10. Review the results.

Note

Check the error percentages for the expired and inspired
volumes. Both volumes must be less than ±3.5% for the
calibration to be acceptable. For single-flow calibrations, the
measured and adjusted curves should match.

Note

The syringe used to check the volume calibration of
spirometers must have an accuracy of 15 mL for a 3-L
syringe.

11. Touch the Accept button to save the calibration results.

Directions for use

About calibration 23