About pre- and post-testing, About effort replacement – Welch Allyn CP 150 spirometry option - User Manual User Manual

Page 34

Advertising
background image

FEV1 — forced expiratory volume 1, the volume of air that is exhaled at one second

of a forced expiration

About pre- and post-testing

If desired, a spirometry test may include both pre- and post-efforts to measure the

effectiveness of medication. The “before medication” and “after medication” efforts

may be uninterrupted or interrupted.

Uninterrupted — If there is no interruption between pre- and post-efforts (that is, no

other patient has been tested and the device has remained on), the same screen

continues to display. You simply continue with the procedure.

Interrupted — If there is an interruption (that is, another patient has been tested or

the device has been turned off), you need to recall the patient’s test-in-progress

before continuing.

Note

Pre- and post-efforts must happen on the same day. The next day tests
become available for review only; you can no longer add efforts to them.

About effort replacement

You can save up to 6 FVC efforts per test. After saving 6 efforts of a given type, the

software compares each new effort with the saved efforts. If the new effort is better

than the worst saved effort, the worst effort is deleted and the new one is saved. If the

new effort is worse than all saved efforts, you are asked whether you want to save it.

If 6 pre-efforts have been saved, the worst pre-effort is deleted when you add a post-

effort until you have saved 3 pre- and 3 post-efforts. After that, the “worst” post-effort is

deleted.

Performing a new Forced Vital Capacity spirometry test

CAUTION Patient data is not saved until the spirometry test is completed.

Note

The spirometry configuration settings can be changed in the Settings tab.
The following settings may appear differently if the default settings have
been modified.

1. From the ECG Home screen touch the

(Spirometry) button.

Note

If the Daily Reminder setting is enabled, the first time this
button is pressed each day, the prompt “calibrate now?”
appears.

2. Touch the

(Perform new Forced Vitals Capacity test) button. The Summary

tab appears.

3. Enter the following patient information:

Note

Required fields are denoted with an asterisk.

Patient ID*. Touch the OK button.

Birth date*. Touch the OK button.

Last name*. Touch the OK button.

30 Spirometry tests

CP 150 spirometry option

Advertising