Note

Obtain the transducer lot and calibration codes from the
transducer package label.

Note

For the syringe volume, see the sticker on the calibration
syringe.

Note

Humidity (%), Temperature, and Pressure are set through
the USB spirometer sensor and are not editable fields. The
temperature must be 10°– 40° C, 50°–104° F. The
atmospheric pressure must be 600 –1100 mbar, 450 – 825
mmHg, 18 – 32 inHg, 60 – 110 kPa.

4. Touch the (Next) button.
5. Connect a new flow transducer to the pressure tubing.
6. Attach the flow transducer to the syringe’s port, as shown in the illustration. Push

the flow transducer all the way in for a tight seal.

7. Touch the Continue button.

CAUTION Several things may affect calibration results:
movement of the syringe, movement of the pressure tubing,
or blockage of air. Place the syringe on a hard, level surface
with at least 1 cubic meter of open air surrounding the flow
transducer. Place your hand on top of the syringe to prevent
movement.

8. Touch the Start button to begin the calibration.
9. When the black bar begins to move, push the plunger all the way in, then pull it all

the way out, carefully following the black bar’s rate. Use a steady motion in both

directions.

22 About calibration