Cardioperfect workstation, Spiroperfect module – user manual – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual

Page 23

Advertising
background image

CardioPerfect Workstation

SpiroPerfect Module

– User Manual

DIR 80012334 Ver. F

23 / 95

Figure 5.3 Spirometry Viewing tab



Setting

Description

Parameters in

Columns or Rows. Changes the layout of the six parameter table.

Show
Predictive

Points and/or Curve.
If Points is checked, predictive points display and print in the FVC graph,
Predictive points definition see page 93.
If Curve is selected a prediction curve will be displayed in the FVC graph.

Trending

% Reference Value or % Predicted. When % Reference Value is selected,
parameters values are graphed as a percentage of the selected reference
value. When % predicted is selected, parameters will trend as a
percentage of predictive values.

Manual
selection of
best effort

If checked, you are allowed to manually select the best effort, when the
Final Result is set to Best Effort.

Display ATS
Acceptability
per effort

If checked, a row or column appears in the Parameter and Measurement
tables displaying whether or not each individual effort meets the ATS 2005
acceptability criteria.

Superimpose
FV Curves

If checked, curves are offset on the graph. If unchecked, all curves are
superimposed.

Advertising