Figure 7.5 Accept stroke
The following options apply:
Yes:
Continue to the next stroke, or show calibration results.
No:
Redo the current stroke with the same speed.
Cancel: Stop calibrating. The sensor will not be calibrated.
9. If the calibration was successful this will be displayed. You can either accept the
results or recalibrate. If the ATS standard has not been met you have to recalibrate.
See the following section on how to deal with the results.
If you have pressed the Recalibrate button: Press the Repeat Calibration button in
the Calibration window and follow the instructions above from step 6.
Note: If you have trouble getting the results within the Calibration Error Tolerance try:
waiting 1 second between emptying and filling the syringe.
a single flow calibration before the three flow calibration, if the three flow
calibration is unsuccessful