Cardioperfect workstation, Spiroperfect module – user manual, Warning – Welch Allyn CardioPerfect Workstation SpiroPerfect Module - User Manual User Manual

Page 34

Advertising
background image

CardioPerfect Workstation

SpiroPerfect Module

– User Manual

DIR 80012334 Ver. F

34 / 95

Setting

Description

Calibration
Error Tolerance

Select the measured calibration accuracy to be within 1% or 3% of the
syringe volume.

Tip: See page 28 on how to change the default setting.

Temperature

The ambient temperature
See Operating Environment Specifications on page 80

Humidity

The ambient humidity
See Operating Environment Specifications on page 80

Pressure

The ambient pressure
See Operating Environment Specifications on page 80


WARNING

For the ambient settings pressure field please enter the
pressure as given by a barometer in the immediate vicinity.

Do not enter the normalized sea-level pressure as commonly
listed on internet sites on meteorological data resources.

2. Press the Next button to continue.


The following screen appears depending on your setup:

Figure 7.2 Calibration window


Beneath the Calibration window the calibration instruction and message window is
displayed, giving instruction for the calibration procedure.


Note: Please check the calibration and lot code if you cannot continue to the
Calibration window.

3. Connect the syringe to the new flow transducer.

4. Fill the syringe by pulling the plunger completely out.

5. Press the Calibrate button in the window.

6.

Wait until the messages ‘Initializing sensor. Opening sensor, please wait...
disappears.