Electromagnetic Immunity

The Welch Allyn SpiroPerfect is intended for use in the electromagnetic environment specified below. The customer

or the user of the Spirometer should assure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance level

Electromagnetic environment

guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

6 kV contact

8 kV air

6 kV contact

8 kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.

Electrical fast
transient/burst

IEC 61000-4-4

±

2 kV for power supply

lines


±

1 kV for input/output lines

Not Applicable

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

1 kV differential mode

2 kV common mode

Not Applicable

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage dips, short
interruptions and
voltage variations on
power supply input lines

IEC 61000-4-11

<5 %

U

T

(>95 % dip in

U

T

)

for 0,5 cycle

40 %

U

T

(60 % dip in

U

T

)

for 5 cycles

70 %

U

T

(30 % dip in

U

T

)

for 25 cycles

<5 %

U

T

(>95 % dip in

U

T

)

for 5 sec

Not Applicable

Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Welch Allyn SpiroPerfect
requires continued operation
during power mains
interruptions, it is recommended
that the Welch Allyn SpiroPerfect
be powered from an
uninterruptible power supply or a
battery.

Power frequency (50/60
Hz) magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

NOTE

U

T

is the AC mains voltage prior to application of the test level.