Propaq 6.2X-8.4X/Propaq Encore 1.20-1.3X Reference Guide

7

Update

Change Other Power-On Settings

To change all other power-on settings such as ECG size, alarm
limits, or printer settings:

1.

Select the desired monitor settings
(for example, select ALARM TONE as HIGH).

2.

From the Main Menu, press:

Propaq Encore:

SETUP, MORE, MORE, PROGRAM, CURRENT, YES

.

Propaq 100 Series:

SYSTEM, (MORE), PROGRAM, CURRENT

.

The monitor selects all of the current settings as the power-on
default settings.

WARNING

If any alarms are set to OFF and you select CURRENT (to set current
settings as power-on settings), those alarms will be OFF at power-on for
subsequent patients. Consider carefully before setting power-on alarms
to OFF.

note...

The alarm for apnea cannot be turned off at any time.

note...

If you change the patient mode, the alarm limit settings are automatically
changed to the defaults for that mode.