5. Place an empty filter bag in the Bag Holder in an open position.

6. Tare the weight of the empty filter bag and the holder together.

7. Add 0.95 – 1.00g of sample to the filter bag. Keep all particles away from the

sealing area of the filter bag.

8. Record the weight of the sample (W

2

) and tare.

9. Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red

sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of
the filter bag. If the seal is not strong, re-seal the bag.

10. Spread the sample out uniformly within the filter bag by shaking and flicking the bag to eliminate

clumping.

11. Repeat steps 5 – 10 for all filter bags that will be used in the Analyzer. (Up to 24 bags can be processed

during one procedure.)

IMPORTANT

For all samples you will need to do a pre-extraction of fat before doing a Crude Fiber
analysis in the ANKOM

2000

. Follow the pre-extraction steps below:

1. Place the filter bags with sample into a 250ml container.
2. Pour enough petroleum ether into the container to cover the bags.
3. Allow the bags to soak for 10 minutes.
4. Pour out and dispose of the petroleum ether.
5. Place the bags on a wire screen to air-dry.

12. Place the filter bags with sample and at least one empty bag (used as a Blank) into

the Bag Suspender trays as shown (maximum of three bags per tray).

13. Stack each tray on the Bag Suspender rod (eight trays in total) with each tray

rotated 120 degrees from the tray below.

IMPORTANT

You must use all eight trays even if they are empty.

14. Add the 9

th

tray to the top of the Bag Suspender rod. This tray contains no filter bags and acts as a cover.

The samples are now ready for the Crude Fiber analysis procedure.