Periodic Maintenance

Initial Maintenance (after 10 hours of operation)

After the first 10 hours of operation, follow the procedure below:

1. Remove the back panel of the instrument.

2. Inspect the Maintenance Alert Collector.

Rear View of instrument with cabinet back removed

3. Clean any residue from the collector.

4. Unscrew the filter body (the fixture on the

outside of the instrument that attaches to the
water supply).

5. Unscrew and clean the metal filter screen.

6. Reassemble the filter and compression fitting.

Make sure the valve body is tight and the
Teflon washer has not come out of the filter
body.

IMPORTANT

Do NOT insert the Bag Suspender into the instrument for this procedure.

7. Press the arrow keys on the Keypad until you see “Select Flush” on the

Display. Press “ENTER” on the Keypad to turn the agitator on.

Select Flush

 v <Enter>

8. When the motor is activated, turn the Packing

Nut to the RIGHT until you hear a change in
the sound of the motor. (The motor will start to
labor as the packing nut gets harder to turn.)

9. Loosen the Packing Nut slightly until the

motor stops laboring.

10. Turn off the instrument and re-install the back

panel.

Packing Nut

Maintenance Alert Collector