Operator’s Manual


pg. 34

Rev E 1/14/15

If you see a leak

If you see a leak, follow steps 6 – 9 in the Initial Maintenance procedure above.

Replacing the Fuses

To replace the fuses in the ANKOM

2000

Fiber Analyzer, follow the procedure detailed below.

1. Turn off the instrument power and unplug the

Power Cord from the outlet.

2. Using a flat blade screwdriver, twist the slot on

the fuse holders counterclockwise ¼ turn to
open.

3. Check both fuses.

4. Pull the fuse from the grey fuse cap and

replace as needed.

(120V – 15 amp / 220V – 10 amp)

Cleaning the Fiber Optic Level Sensor

Using a cotton swab with alcohol, wipe the tip of the
level sensor at least once per month if you use the
instrument daily.

Level Sensor

Fuses (2)