5 Putting into operation

26

K-415 Operation Manual, Version B

5 .1 .1

Anti-seismic-tie-down

For the use in earthquake-susceptible regions, the instrument can be equipped and secured with the
anti-seismic-tie-down, contained in the standard delivery of the instrument.

The fixing device can be attached to the instrument from the bottom side of the instrument, close to
the housing foot on the right rear side of the instrument.
• Slide the fixing bracket a into the provided groove b and fix it with the provided screw c.
• The instument can now be tied to the setup.

Fig. 5.1: Fixing the anti-seismic-tie-down

5 .2

General installation procedure for all Scrubber models

To commission the Scrubber K-415, proceed as follows:
• Make sure that the silencer (a in figure 5.2) is filled with polyester fibres and install it:

1

Fig. 5.2: The silencer on the rear side of the instrument

• Check the glass for flaws and cracks.
• Check the seals and gaskets (embrittlement, scratches). If they are not OK, take them out and

replace them (see also chapter 7).

• Carefully insert the glass parts. Otherwise there is a risk of damaging the glassware.