2 hose connections for the triplescrub – BUCHI Scrubber K-415 User Manual

Page 33

Advertising
background image

5 Putting into operation

33

K-415 Operation Manual, Version B

5 .7 .2

Hose connections for the TripleScrub

The TripleScrub consists of the neutralization vessel, the adsorption vessel, and the condenser unit.
To establish the hose connections, proceed as follows:
• Put hose connectors GL 14 with seals on all PTFE hoses.
• Connect the emission source to the condenser inlet

e

.

• Connect the condenser outlet

f

to the neutralization vessel

a

.

• Connect the adsorption vessel

b

to the bypass valve

c

.

• Cut off a piece of the provided silicone hose and connect the cooling water source to the cooling

water inlet of the condenser

g

. Secure it with a hose clamp.

• Cut off a piece of the provided silicone hose and connect the cooling water outlet of the condenser

h

to the waste water system.

1

2

3

4

5 6 7

8

9

a Neutralization stage
b Adsorption vessel
c Bypass valve
d Silencer
e Inlet for fumes from emission source

f Outlet to neutralization vessel (top)
g Cooling water inlet (bottom)
h Cooling water outlet (top)
i Condenser

Fig. 5.12: Hose connections for the TripleScrub

Advertising