6 Chapter

2

Front Panel (optional)

Monochrome 480 x 272 LCD display with backlight

Five tactile keys for menu item selection

Load Cell Excitation

5 VDC +/- 1.15 VDC maximum

Isolation from digital section 1000 VDC minimum

C2 Calibration Input

Isolation from digital section 1000 VDC minimum

Cable Lengths

250 feet maximum of C2 authorized cable (Maximum of 4 load sensors) with IT

Junction box

Network Connectivity

Ethernet TCP/IP

Ethernet UDP

Modbus-RTU over RS485

HI 6310 Only
Analog 4-20 mA

Printer port

USB port

Environmental Requirements

Operating Temperature Range

0 to 60º C (32º to 140º F)

Temperature Coefficient

Less than 0.005% of full scale per degree C for Cal-LO and Cal-HI reference points

Storage Temperature Range

-40 to 85º C (-40º to 185º F)

Humidity Range

0-90% (non-condensing)

Maximum Altitude for Installation

2000 Meters (6,562 Feet)

Pending Approvals

UL, CUL and CE

Hazardous Class I, Division 2, Groups A,B,C,D, T4A and Class I, II, III, Division 2,

Groups F, G, T4A