IDS 430/433 Users Manual Version 1.0

- 57 -

Diagnostic Tests

The DIAGNOSTIC TESTS are accessed using configuration function 69. Hold

the CLEAR key down and press the ENTER key. The display prompts “CFG xx”

where xx is the currently selected parameter. Enter 69 and press the ENTER key

to access the DIAGNOSTIC TESTS.
The display prompts “diA xx” where xx is the last selected diagnostic function.

Select a test number from the table below. Enter the test number and press the

ENTER key.

Table of Diagnostic Tests

Test No.

Hardware Tested

Description

1

Serial Com. Port 1

Display input data.

2

Serial Com. Port 1

Display input error count.

3

Serial Com. Port 1

Transmit data.

4

Serial Com. Port 2

Display input data.

5

Serial Com. Port 2

Display input error count.

6

Serial Com. Port 2

Transmit data.

7

Parallel Port

Transmit data.

8

A/D converter

Display raw conversion data.

9

Ram Memory

Test memory for errors.

10

Display

Turn on all LED’s and display segments.

11

Configuration

Print the EAROM configuration data.

12

Serial Com. Port 1

Transmit/Receive loop-back test.

13

Serial Com. Port 2

Transmit/Receive loop-back test.

Diagnostic Test 1: Serial Com Port 1 - Display Input Data

This test displays serial data as it is received by serial communications port 1.

The numeric display has limited alpha display capability, however numeric and

some alpha characters are legible. If all data being received is unintelligible, the

baud rate is probably incorrect. If nothing is displayed on the display, then check

the following:

1.

RS232/Current loop switch on the main circuit board.

2.

Cabling between the communicating units.

3.

Make sure the sending unit is actually sending data.

Press the CLEAR key to exit diagnostic test 1.