C5 gardner colour number – ALTANA LCM plus User Manual

Page 13

Advertising
background image

BYK-Gardner GmbH

LCM plus

235 017 133 DE 0302

42

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment. Measurement starts auto-

matically.

2. After the measurement, the colour value is indicated on the

display.

3. Now press the

-key (DIM) until the Hazen number appears on the display.

C5

Gardner Colour Number

The Gardner colour number is defined in DIN ISO 4630. The lighter Gardner colour standards (1 to

8) are based on potassium-chloroplatinate solutions, colour numbers 9 to 18 on ferric(III)-chloride,

cobaltous(II)-chloride and hydrochloric acid solutions. LCM plus determines the Gardner number at

all cuvette path lengths (in 50mm cuvettes only values between 0 and 4), with 11mm cuvettes

providing sufficient accuracy.

Measuring procedure

1.

Fill the sample liquid into a cuvette and insert it into the

corresponding compartment.

2.

After the measurement, the colour value is indicated on the

display.

3.

Now press the

-key (DIM) until the Gardner number

appears on the display.

C6

Mineral Oil Colour Number ASTM D 1500 And ISO 2049

The Mineral Oil Colour Scale is used for matching the colours of mineral oil products like lubricating

oil, fuel oil, diesel oil and paraffins. The colour scale begins at 0 for water-clear, colourless oils

and ends at 8 for very dark brown oils. Visual matching is in steps of 0.5 (0.5 1.0 1.5...). LCM

displays the results in steps of 0.1. For samples of stronger hue than colour glass 8, colour value

D8 is displayed.

Owing to the intensive colour gradation of the ASTM

9

D 1500 colour number, this is only

determined for the 11mm round cuvette. There is no determination of path length 50mm.

9

American Society for Testing and Materials

20

Hazen color value

_DIM ¢Conf 50°

6.5

Gardner

color

value

_DIM ¢Conf
11°

Advertising