D2

Maintenande and Test Media Control

Even a perfected product like LCM plus requires careful handling. Use a dry cloth or a mild

household cleaner for the housing, but never solvents like naphta, acetone or the like. The optical

elements should be cleaned by our maintenance staff.

A certified set of test solutions is available for test media control according to ISO 9000ff with

order no. 6384.

The test set comprises six bottles

containing 50ml of test solution each,

a quality control certificate according

to DIN 55350 part 18, 4.1.2 and a test

protocol for the documentation of

reference,

actual

and

tolerance

values.

LCM plus is tested as described in

enclosed test instruction.

The

instrument should be tested in regular

intervals of 1 to 2 weeks, depending

on its application and frequency of use.