Ccolour measurement, C1 calibration – ALTANA LCM plus User Manual

Page 9

Advertising
background image

BYK-Gardner GmbH

LCM plus

235 017 133 DE 0302

38

C

Colour Measurement

Whenever the instrument is turned on, a zero measurement is made, taking the empty value (i.e.

the pure transmission values without cuvette inserted) of the optical beam path and comparing it

with the empty value obtained in the previous zero measurement. If the difference between both

empty values is near zero, you can start measuring without calibration. Calibration is only required

in case of considerable deviations of the empty value due to e.g. fluctuations of temperature. The

calibration remains valid until the next recalibration. LCM plus features a calibration memory for

both path lengths employed (50mm square cuvette and 11mm round cuvette). If you have to use

two different path lengths, owing to the nature of your samples (e.g. bright samples with Hazen

values below 50 or strongly coloured samples with greater Gardner values), it is possible to

calibrate both path lengths and then toggle between them. LCM plus uses the corresponding set

of calibration data automatically.

C1

Calibration

LCM plus features a calibration memory for 11mm Dr. Lange round cuvettes and 50mm square

cuvettes. To be on the safe side, calibrate LCM plus at least every day. In the calibration, the

employed medium is assigned a transmission of 100% over the whole spectral range. Observe

the following rules for calibration:

The cuvette type selected for calibration must be identical to the one used for

measurement, otherwise wrong Iodine, Hazen and Gardner values are given.

Use always clean cuvettes and fresh distilled water for calibration.

Instrument calibration for colour measurement

2.

Turn-on instrument and printer, if desired. After self-check

and zero measurement, LCM plus might ask for calibration.

If LCM plus is already on, make sure the cuvette

compartment is empty and press the blue

CAL

- key. The

display reads:

2.

Fill a clean cuvette (material and path length to suit your

samples) with distilled water and put it into the measuring

compartment.

4.

LCM plus detects the cuvette type automatically.

4.

Now press the blue

CAL

- key again to trigger calibration.

5.

When the calibration is over and the printer is on, date, time and selected cuvette type are

printed, e.g. :

Zero meas.m.

Please wait

Pleas ins. cuv.

for calibration

Calibration

Please wait