Lcm plus, B5 turning-off, B6 photometric preparation – ALTANA LCM plus User Manual

Page 8

Advertising
background image

BYK-Gardner GmbH

LCM plus

235 017 133 DE 0302

37

B5

Turning-Off

LCM plus turns off automatically by a timer function. Manual turning-off is possible, too.

Manual Turning-Off

Press Cursor-Key

in the measuring menu and select option

Switch Off“ with the OK-key. Then the instrument turns off.

Turning-Off Via Timer

On delivery, the timer of LCM plus is set to automatical turning-

off time of 30 minutes. The automatical turning-off is only

active if no measurement was made within the given period

and no cuvette is in the compartment. The setting can be

modified within the tolerances of 10 to 60 minutes (cp. C8.2).

B6

Photometric Preparation

The photometric preparation requires utmost care, because careless sample preparation is the

major source of trouble, and not the handling of the instrument itself.

Cleanliness is of vital importance in photometry. Therefore, clean the glass cuvettes immediately

after use. Employ only optically perfect samples for measuring. They must be clean and not show

any signs of opaqueness.

Air bubbles in the sample or at the cuvette sides influence the measuring result. They can be

avoided by filling the liquid slowly into the cuvette. Pasty or solid products should be melted and

filled-in without bubbles, if they cannot be measured as they are.

B6.1

Sampling And Sample Preparation

Take a sample from the product you want to measure in accordance with DIN 53242 part 1.

Eliminate turbidities by heating or filtration. Heat partly solid samples before measuring in order to

dissolve the solid share in the liquid one. The preparation must not cause any chemical changes in

the sample. During the measurement there should be no bubbles in the sample.

6

There are several cuvette types available for colour measurement with LCM plus,

differing by material (glass, polystyrene or PMMA

7

) and path length (11mm and 50mm).

The program calculates automatically Iodine, Hazen, Gardner, Saybolt and ASTM D 1500-

numbers under consideration of the cuvette type employed.

For the 11mm disposable round glass cuvettes, a dry thermostat is available to warm up

to 15 cuvettes to 100°C or 148°C.

6

Acc- to DIN 6162, ed. 12.81, para. 5 a. 6

7

Polymethylmethacrylate

LCM plus

V1.05