D3 lamp exchange, D4 rs232c-interface of lcm plus – ALTANA LCM plus User Manual

Page 19

Advertising
background image

BYK-Gardner GmbH

LCM plus

235 017 133 DE 0302

48

D3

Lamp Exchange

The lamp compartment is left of the display and covered with a slotted cap on the upper side of the

housing. If the lamp is broken, it should be exchanged by Dr. Lange service staff, because w e

recommend a general instrument check-up after an operating period of 1-2 years. If you insist on

doing it yourself, proceed as follows:

R e m o v e p o w e r s u p p l y u n i t o r b a t t e r y!

Remove the grey cover of the bit-box. Loosen

the fixture of the bit-box by pushing the locking

clamp

and put it apart. Unplug contact

from the lamp. Loosen the metal clamps

at

the upper end one be one from their stay and

hinge them away. Take the lamp

off and put

in a new one with the marking pointing to the

left.

Attention: Do not touch the glass bulb of

the lamp.

Hinge the metal clamps back to their places and

fix them to their stay. Connect the new lamp

with the plug-in contact

and do the same

steps as for dismantling, but in reverse sense.

Only

when

the

instrument

is

mounted

completely, turn it on and test it.

D4

RS232C-Interface of LCM plus

LCM plus has a serial interface at its rear side (page 7, Abb. 3) corresponding to RS 232 C and

V24-standard.

Fig. 15: Removing the fixture of the
bit-box

Fig. 16: Lamp exchange

Advertising