6.1 Cyan Check

Choose "Cyan Check" from the "Calibrate" menu.

Open the cover of the cyan checking standard and
position the instrument on the standard. The
instrument snaps into the indentation of the
standard and can only be positioned in one
direction.

Choose "OK" with the scroll wheel.

When the cyan check is completed, an affirmation
is shown in the display.

Confirm by pressing the operate button or the wheel
to return to the main menu.

If the reading of the cyan standard was out of
tolerance, a black and white calibration is
necessary. The instrument leads you directly to the
calibration display.