Maintenance and repair – ALTANA Byko-Drive Manual User Manual

Page 16

Advertising
background image

14

English

Maintenance and Repair

6. Maintenance and Repair

Cleaning and Routine Maintenance
Care should be taken to avoid letting coating
materials into the vacuum plate holes. If cleaning
is needed the holes can be cleaned with a solvent
appropriate for the coating material and a wooden
toothpick. Metal tools should be avoided as they
may damage the aluminum vacuum plate.
Clean the rest of the instrument with a mild com-
mercial cleaning solution or plain soapy water. Avoid
solvents on the cover and the keyboard area as they
may be discolored or otherwise damaged by strong
solvents.

Troubleshooting
If the Pushbar becomes obstructed during move-
ment it will stop and the Status LED will turn red.
Turn off the Power Switch and remove the obstruc-
tion. Turn the Power Switch back on and the byko-
drive will reset to the home position.
If the Pushbar will not move when either the For-
ward or Reverse buttons are pushed and the LED
remains green, the start position and end positions
may have been mistakenly set in the same place.
Simply push and hold the Set Start Position button
for more than 5 second and release, the LED will
blink when the button is released. Repeat this step
for the Set End Position button. This will clear the
set points so the Forward and Reverse button again
function as expected.

Service and Spare Parts
For all service and spare parts requirements, please
contact your local BYK-Gardner office.