14

-------Prior to initiating a totally different work procedure

-------When the safety office or safety committee requested

NOTE: Only the safety office and the safety committee can determine the correct

sterilization procedure.

For the convenience of the vapor sterilization, the following cabinet size is available (see chart
7):

(Chart 7)

MODEL

OVERALL SIZE

WORKING AREA SIZE

Weight

L x W x H

Inch (mm)

L x W x H

Inch (mm)

Lbs (Kg)

11228-001

55.59 x 33.46 x 85.82

(1412 X 850 X 2180)

85.82 x 26.18 x 25.59

(1220X665X650)

838

(380)

TESTING AND CERTIFICATION

The ENVIRCO Biohazard Cabinet has been fully tested by the manufacturer prior to

shipment; however, shipping and handling during installation may result in leaks in the HEPA

filter or other cabinet seals. The cabinet should be adjusted for proper airflow balance after it

has been installed in the final location.

Recommended Tests

The following tests are recommended at initial installation:

1. Aerosol Leak Test of the recirculating air HEPA filters

2. Aerosol Leak Test of the exhaust air HEPA filter

3. Adjustment of the airflow within the cabinet to an average velocity per values on page 10 of

this manual

4. Adjustment of the exhaust air volume to provide a minimum inward airflow per values on

page 10.

5. Visually verify that an inward airflow exists around the entire periphery of the face opening

using an acceptable smoke challenge

6. Check electrical continuity between cap plug ground pin and exposed metal cabinet

surfaces to verify that electrical resistance does not exceed 0.15 ohms. Initial on-site testing

and certification of the ENVIRCO Biohazard Cabinet is included with the purchase of the unit.

Contact an AuthorizedService Contractor to arrange for certification date. Testing and

certification service includes operation and maintenance training for the unit and validation of

the manufacturer’s warranty. Operation of the cabinet prior to authorized certification may

void the warranty and service contract or result inadditional service contractor charges.

Authorized Service Contractor: ENV Services, Inc. (800) 345-6094Service Record