Hard connection is only used in special applications with type A2 Biological Safety Cabinets.

Should your situation require such a configuration, contact Envirco for information.

6.3 Instructions for flow alarm monitor system

Flow alarm monitor system is a pre-programmed, self-contained package that monitors

differential static air pressure. The alarm would note when the air pressure is ±20% beyond

the normal pressure. When the air pressure is 20 % lower than the normal pressure, it hums

three times every five seconds, and when the air pressure is 20 % higher, four times every five

seconds.

FIGURE 11: EXHAUST SYSTEM