Introduction, I、 introduction – Envirco BSC User Manual

Page 4

Advertising
background image

4

I

Introduction

Investigators and technicians concerned with microbiological safety have, for many years,

utilized specialized containment enclosures to limit their exposure to harmful pathogens. In

addition, it is recognized that exacting research, as well as routine pathology, is often

compromised by environmental contamination. The ENVIRCO Class II Biohazard Cabinet is

designed to limit the contamination exposure of both the worker and the work when handling

biohazard material. The Class II enclosure is an open-face cabinet with a high volume of air

circulated through an internal HEPA filter. Air is drawn into the open front of the cabinet and

exhausted through a HEPA filter. The controlled airflow direction and volume balance provide

protection for the worker, the work (sample), and against cross-contamination within the

cabinet.
This product has been tested to meet the requirements of CAN/CSA-C22.2 No.61010-1,
secondedition, including Amendment 1,or a later version of the same standard incorporating
the same level of testing requirements.

Important

Volatile, friable or poisonousradioactive substancesarenot allowed in the

non-exhaust-cover cabinet.

 Sterilize the cabinet before turning off PAO test switch or wash water valve

 Do not hinder airflow flowing out of HEPA filter.

 Disconnect power before accessingthe operational board.

 Maximum power of this equipment is 1000W.

 Connect Only To Properly Grounded Outlet. Do Not Remove Ground Pin.

 If the equipment is used in a manner not specified by the manufacturer, the

protection provided by the equipment may be impaired.

 If needed, please use the emergency power cut-off, located on the front panel.