5.4Fault finding hints

Problem Solution

The equipment does not

function

(a) Check power at the electrical outlet into which the

equipment is plugged.

(b) Checkthe following three fuse tubes on the circuit to

determine if they are broken:Zero wire fuse tube F4

12.5A, Fire wire fuse tube F5 12.5A, Transformer fuse

tube F6 630mA

(c) Transformer failure or poor connection with wires cannot

provide normal working voltage

Blower does not function

(a) Check the motor

(b) Make sure door is completely closed for UV

(c) Check control panel wiring

(d) Check to determineif fuse tube of blower is broken

Motor hums

(a) Replace the motor

(b) Fan speed is set too low

Lights do not

function(fluorescent or

U.V)

(a) The lamp is broken

(b) Some problems with the units on operational panel or

wires.

(c) Replace ballast

(d) Check to determineif the fuse tube of light(F1 2.5A) and

UV lamp( F2 2.5A) is broken

Rattling or scraping noise
from sash motor/blower

(a) Check for debris

、paper、wrappings、etc., and clean up

(b) The blower may have slipped on the motor drive shaft;

replace the blower.

(c) Check the emergency sash stop, ensure it is not pressed

down

Air flow velocity too low

(a) Fan speed is incorrectly set

(b) HEPA filters are loaded; check the static pressure monitor

reading.

(c) Supply/exhaust air flow requires balancing.

Glass door motor not

working

(a) Check to determine if the fuse for the motor is blown,

F8 6.3A

(b) Check motor; replace the motor if necessary