4 totalization, Totalization – KERN & SOHN VFS 2T1 User Manual

Page 24

Advertising
background image

VFS-BA-e-1111

24

7.4 Totalization

The display unit offers an option that allows the addition of weighing values and the
display of the total weight. If a tare weight is active, the net weight is automatically
added up.

Ö No load may be on the loading fork
Ö The system should be stable (stability display „STABLE

illuminated)

Ö The added-up weighing value must be more than 20d.


Ö Loading the system with the load to be added

Ö Wait for stability display „STABLE“, then press

, the weight value will be

added into a sum memory

Ö „ACC 01“ will be shortly displayed, after that the system changes into the

standard weighing mode. Next to the weight display „M+“ appears

Ö Unload the system, the display returns to „0.0
Ö 2. Place the weighing good, wait for stability display „STABLE

Ö Press

anew

Ö „ACC 02“ is displayed, after that the total sum of the weighing procedures will be

displayed briefly before the system returns to weighing mode

Ö Carry out further weighings in the same way. Ensure that the system is unloaded

between the weighing procedures.

Ö The weighing system maximally can carry out 99 single weighing procedures

Ö To print out data unload weighing system and press


Printout example (see chap. 8 Printer):
Date 2011/07/04 Date

Time 14:03:26 Time

No.

3 Number of single weighings

N.W.

30.0kg Weight of the last single weighing

TOTAL 66.0kg Total

weight




Advertising