 Press

for approx. 3 sec., the count down for setting

the second adjustment point runs off automatically.
The weight for the second adjustment point is displayed.
Moreover two points appear in the display.

 Place 2000 kg

 After placing the weight the weighing system leaves the

reference height, the display flashes

 Place the weighing system on reference height, the display

stops flashing.

 Press

for approx. 3 sec., the count down for setting

the third adjustment point runs off automatically, 0 is
displayed

 Remove weight.

 Press

, vU004 is displayed, followed by PAr

 Press

for approx. 3 sec., SAVE is displayed,

followed by 0.
The balance is now in weighing mode.

Adjustment now has finished.

An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.