Carry out adjustment by entering the adjustment weight

 Switch off weighing system

 Press

for 30 sec., PAR is displayed

 Press

until CAL is displayed

 Press

, ZERO is displayed flashing, because the

reference height is not reached yet

 Place a pallet and adjust the reference height, when the

reference height is reached, the display stops flashing and
only the righthand bar above in the display continues to be
displayed.

 Press

for approx. 3 sec., the count down runs off

automatically.


.1000 is displayed.

 Press 2 x

, 0 is displayed

 Press

shortly, the display for entering the

adjustment weight is displayed, the right digit flashes.

 Use both keys to enter the desired adjustment weight:

Use

to alter cipher, use

to change the digit

(example)

INFO: when entering 1500 kg = enter 15000 (example). With the other values
proceed in the same way: the last zero will not be considered: