11 Instant help


In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.


Help:

Fault

Possible cause

The displayed weight does not glow.

• The balance is not switched on.

• The batteries are empty

The displayed weight is permanently
changing

• Draught/air movement

-

• Sites with vibration.

• Loading fork is in contact with foreign

matter.

• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing value is obviously

• The display of the balance is not at zero

wrong

• Adjustment is no longer correct.

• Great fluctuations in temperature.
• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

• Load fork has not been pumped to

reference height

• The control pin is not in position

„Reference height“.


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.