5.6

Weighing

Load the suspended balance.
The weight value will be displayed at once.

Overload warning

Overloading exceeding the stated maximum load (max) of the balance, minus a possibly
existing tare load, must be strictly avoided. This could cause damage to the balance.
Exceeding the maximum load is indicated by the display „ol“ . Unload balance or reduce
preload.

5.7

Switch-over weighing unit


The next weighing unit kg

Nlb will be displayed after each press of the UNIT

button.
The indicator

 above N shows weight in Newton.



In menu item “F2 UNt“, weighing units “N“ and “lb“ can be turn on or off.