Adjustment – KERN & SOHN HFA 10T-3 User Manual

Page 32

Advertising
background image

7. Adjustment

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.

• Provide the required adjustment weight, see chap. 1. “Techn. data“.

The weight to be used depends on the capacity of the scale. Carry out
adjustment as near to maximum load as possible. Info about test weights can
be found on the Internet at: http://www.kern-sohn.com

• Observe stable environmental conditions. A warming up time (see chapter 1)

is required for stabilization.

Turn off balance and attach suitable lifting tackle.

Turn on balance with the lifting tackle attached and press

during self-test. The first function „F0bk“ is displayed.

Press

repeatedly until „F5CAL“ is displayed.

Press

, the last parameter set will be displayed.

• „nonLi“

Adjustment of balance

or

• „LinE“

Linearisation of balance

Select „nonLi“ by

32

HFA-BA-e-1511

Advertising