Attach second adjustment weight, wait for stability display

Press

, “LoAd 3“ (max load) will be displayed

Attach third adjustment weight, wait for stability display

Press

, “PASS“ will be displayed

Finally the balance will carry out a self-test, “Err19“ is briefly
displayed (Ignore error message), then the balance returns
automatically to weighing mode, the weight value is
displayed and linearization has been completed
successfully.

(example)

If an error occurs, turn off on restart balance and repeat linearization.