10.3 Peak value function

This function displays the highest load value (peak value) of a weighing.
The peak value remains in the display until it will be deleted.

Mode 1: Menu setting FNC-07 „1“

From a weighing value of >10 d the highest load value will be displayed.

The peak value remains in the display until it will be deleted using

. Then the

balance is ready for further measurements.

Display










Mode 2: Menu setting FNC-07 „2“

From a weighing value of >10 d the highest load value will be displayed. If the weight
value is less than 10d, the displayed peak value will be automatically deleted.

Display