13 installation display unit / platform, 1 weighing system design – KERN & SOHN KXE-TM User Manual

Page 33

Advertising
background image

KXE-TM-BA_IA-e-1411

33

13 Installation display unit / platform

Installation / configuration of the weighing system must be carried out by a
well acquainted specialist with the workings of weighing balances.

13.1 Weighing system design

The display unit is suitable for connection to any analogue platform in compliance
with the required specifications.
The following data must be established before selecting a weighing cell:

Weighing balance capacity

This usually corresponds to the heaviest load to be weighed.

Preload

This corresponds to the total weight of all parts that are to be placed on the
weighing cell such as upper part of platform, weighing pan etc.

Total zero setting range

This is composed of the start-up zero setting range (± 2%) and the zero
setting range available to the user via the ZERO-key (2%). The total zero
setting range equals therefore 4 % of the scale’s capacity.

The addition of weighing scales capacity, preload and the total zero setting
range give the required capacity for the weighing cell.
To avoid overloading of the weighing cell, include an additional safety
margin.

Smallest desired display division

Advertising