14.2 Adjust / Menu item

• The weight to be used depends on the capacity of the scale. Carry out

adjustment as near as possible to the scale’s maximum weight. Info about
test weights can be found on the Internet at: http://www.kern-sohn.com.

• Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.

 Actuate adjustment switch, „SET“ followed by „01 CSP“ is displayed.

 Invoke menu item

using

or

and confirm by

.

The adjustment of the zero point

is displayed.